Used CT scanner inspection checklist for buyers.
Escrow protects your money only if you know what you are accepting. Inspect the CT while it is still installed and scanning, check it again on arrival, and release funds when both checks match the listing.
Short version. Before removal, see the scanner run: photograph the tube usage screen, error logs and installed options, run water and, ideally, ACR phantom scans, and confirm it exports DICOM images. Review the service history. Agree the rigging plan and your site preparation in writing. On delivery, check every component against the inspection record, then release escrow.
Why inspect a used CT before it is de-installed?
A CT scanner can only be tested while it is installed, powered and calibrated. Once the gantry is disconnected and crated, all you can check is its physical condition. So the most useful inspection happens at the seller’s site, before de-installation, with the system scanning. The second check, on delivery, confirms nothing was lost or damaged in between.
Escrow ties this together. Using Escrow.com as the example, the parties set an inspection period of 1 to 30 calendar days, and it starts when the buyer receives the equipment. (Escrow.com) Write into the purchase agreement exactly what acceptance depends on: the pre-removal test results, the arrival check, or testing after installation. Then size the window to fit.
Tube: usage, history and remaining life
- Tube usage screen. Ask for a photo of the console screen that shows tube usage (scan seconds, exposure counts or mAs, depending on the manufacturer). Take your own photo at the inspection.
- Tube identity. Serial number, install date, and whether it is the original tube or a replacement. A recently replaced tube is worth more than one with heavy use.
- Error and arc history. Ask the service engineer to pull tube-related errors and arcing events from the logs. Repeated arcs or aborted scans are a warning sign.
- Warranty. If the tube was replaced recently, ask whether any tube warranty remains and whether it transfers.
The tube is one of the most expensive parts to replace, so remaining tube life belongs in the price. More on why tube data drives value is in selling a used CT scanner.
Gantry, table and detector
- Rotation. Watch and listen to the gantry at its fastest rotation speed. Grinding, knocking or vibration can point to bearing or slip ring wear.
- Covers and cabling. Cracked covers, damaged cables and signs of fluid or pest damage inside the gantry.
- Table. Full travel, height and tilt movements, weight rating and the condition of the tabletop and mattress.
- Detector. Ask for the most recent detector or air calibration date and results. Ring artifacts in phantom images often point to a detector or calibration problem.
- Gantry tilt and lasers. Check tilt range (if fitted) and that the positioning lasers line up with the scan plane.
- Error logs. Ask for the system error log covering recent months, not just the current session.
Service history
- Preventive maintenance records and dates, and who performed them
- Field service reports for major repairs: tube, detector, slip ring, generator, table
- Any outstanding manufacturer field changes or safety notices and whether they were completed
- The most recent annual medical physicist report, if the seller will share it
- Whether the system was on a manufacturer or independent service contract, and whether that service provider can support it at your site
Software level and licenses
Photograph the console’s software version and installed options list during the inspection. The features you are paying for, such as cardiac, perfusion, dose reduction or advanced reconstruction, are usually enabled by license keys tied to the system. Some may need the manufacturer’s involvement to stay active under a new owner, and options on separate workstations may be licensed separately from the scanner. Make the options list part of the purchase agreement and confirm what happens if a license does not transfer.
Also confirm you will receive the service and application software, backup media, manuals and any service keys the system needs to be reinstalled.
DICOM and connectivity
Ask for the scanner’s DICOM Conformance Statement. Under the DICOM standard, the conformance statement lets you determine which optional parts of DICOM the product supports, along with its network roles. (DICOM PS3.2) Give it to your PACS team to confirm the scanner will send images, receive worklists and support any dose reporting you need.
At the inspection, export a test study to media or a network destination to prove DICOM export works, and open it on a separate viewer.
Phantom tests: proving image quality
- Water phantom. Run the daily quality check and record the CT number of water, image noise and uniformity. Look for rings, streaks or bands.
- ACR CT phantom. If one is available, it covers more ground. The ACR describes it as four modules: positioning and alignment, CT number accuracy and slice thickness; low contrast resolution; CT number uniformity; and high contrast (spatial) resolution. (ACR)
- A qualified medical physicist. Having your physicist attend, or review the images and the seller’s latest physicist report, turns a visual check into documented results.
If you plan to accredit the scanner with the ACR, note that the ACR requires phantom exams to have been performed within 14 months of the testing package date, and uses a qualified medical physicist for the Annual System Performance Evaluation or, for new units or units after major repairs, the Acceptance Testing Evaluation. (ACR) Phantom results from the seller’s site are a baseline. Plan for testing again after installation.
De-installation and rigging
- Who de-installs. A qualified engineer should de-install, label and pack the system, ideally the same team or service provider that will reinstall it.
- Packing and transport. Crating or air-ride transport for the gantry, protection for the detector and tube, and climate control in transit.
- Rigging path. Gantry dimensions and weight from the manufacturer’s site planning documents, checked against doors, corridors and elevators at both ends.
- Insurance. Certificates of insurance for the rigger and carrier, and who carries the risk while the scanner is in transit.
- Completeness. A packing list of every component and accessory: gantry, table, generator and power cabinets, console, workstations, injector if included, phantoms, cables and software.
On delivery, check the packing list item by item, photograph the crates before opening and any visible damage, and note it before signing for the shipment.
Site preparation at your facility
- Shielding. A qualified expert should design or review the room shielding. NCRP Report No. 147 sets out structural shielding recommendations for medical x-ray imaging facilities. (Medical Physics summary of NCRP Report 147)
- Power, HVAC and floor. Electrical supply, heat load and floor loading from the manufacturer’s site planning documents for that exact model.
- Radiation paperwork. FDA requires a report of assembly within 15 days after a diagnostic x-ray system is assembled, with copies to the purchaser and, where applicable, the state radiation agency. (FDA) Register the scanner with your state radiation program as it requires.
- Acceptance testing. Schedule your qualified medical physicist to test the system after installation and before the first patient.
When is it safe to release escrow?
- The pre-removal inspection matched the listing: tube usage, options, error logs, phantom results and DICOM export.
- The packing list is complete and nothing arrived damaged.
- Any post-installation conditions in the agreement, such as passing acceptance tests, are met.
- You have the documents: bill of sale, service history, software and manuals, and the seller’s data wipe certificate (see NIST SP 800-88 data wipes).
If something does not match, raise it in writing inside the inspection window. Buying an MRI too? The helium and coil questions are covered in MRI coils and helium.
How TradePark handles this
On TradePark, every seller is a verified business, and your payment goes to a licensed third-party escrow provider, not to the seller, until you have inspected the equipment. You receive one quote covering the scanner, freight and TradePark’s service fee. TradePark coordinates de-installation, rigging and freight with licensed, insured carriers, and the inspection terms you agree are written into the deal. TradePark never takes title, and escrow releases the funds only after you accept.
Start at buy equipment, browse CT scanners and other equipment for sale, or read how a deal works. Selling a scanner instead? List it here.
Questions
When should a buyer inspect a used CT scanner?
Twice. Once at the seller’s site while the scanner is still installed and running, so you can see it scan, read its tube usage and error logs and run phantom tests. And again on delivery, to confirm that what arrived matches what you inspected and was not damaged in rigging or transit. Write both into the purchase agreement as conditions of escrow release.
What tube information should I ask for on a used CT?
A photo of the tube usage screen from the console, the tube’s install date, its serial number, and whether it is the original or a replacement. Ask for the service records for the tube and any tube-related error or arc events in the logs. The tube is one of the most expensive parts to replace, so its remaining life is central to the price.
What phantom tests can be run on a used CT before purchase?
At minimum, the daily water phantom check: CT number of water, image noise and uniformity, plus a look for rings or streaks. If an ACR CT phantom is available, its four modules cover positioning, CT number accuracy and slice thickness, low contrast resolution, uniformity, and high contrast (spatial) resolution. A qualified medical physicist can run and interpret these for you.
Do software options and licenses transfer with a used CT?
Not automatically. Options are usually enabled by license keys tied to the system, and some may need the manufacturer’s involvement to stay active under a new owner. Photograph the console’s installed options list during the inspection, and make the options you are paying for a written condition of the sale.
How long should the escrow inspection window be for a CT scanner?
Long enough to do what the agreement says acceptance depends on. Escrow.com lets the parties set an inspection period of 1 to 30 calendar days, starting when the buyer receives the equipment. If acceptance depends on post-installation testing, the window has to cover installation, calibration and testing; if it depends on a pre-removal inspection plus an arrival check, it can be shorter.
Sources
- Escrow.com: inspection period length (1 to 30 calendar days)
- ACR: overview of the CT phantom (four modules)
- ACR: CT accreditation testing overview (14-month phantom window, medical physicist role)
- DICOM PS3.2: conformance statements
- Medical Physics: summary of NCRP Report 147 shielding design recommendations
- FDA: diagnostic x-ray reports of assembly (15 days)